Results: the pod4 embolization coil was detached from its pusher assembly intact.The diameter of the embolization coil was measured and found to be within specification.Conclusions: evaluation of the returned pod4 confirmed that the embolization coil was detached from its pusher assembly.The complaint cited that no resistance was experienced during advancement or retraction of the coil and that the subject lantern had no visible damage.The pusher assembly and lantern identified in the complaint were not returned for evaluation.Therefore, the root cause of the reported failure could not be determined.Penumbra coils and catheters are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
The patient was undergoing a coil embolization procedure in the splenic artery using a pod4.During the procedure, while attempting to place a pod4 into the target vessel using a lantern delivery microcatheter (lantern), the tip of the pod4 hit the wall of the vessel several times and the pod4 kicked back very hardly.The physician could not control the pod4 in the aneurysm and decided to remove it; however, while removing the pod4 from the lantern, the pod4 unintentionally detached inside the lantern.Therefore, the physician removed the guiding sheath and the lantern containing the detached coil.The procedure was completed using another microcatheter and other coils.There was no noted damage to the lantern after removal.There was no report of an adverse effect to the patient.
|