(b)(4).Fse arrived at the site to address the reported event.Fse replaced the sample needle assembly and adjusted the flow rate & sample uptake.Additionally, fse checked the injection valve seal and stator, performed a drain flush, primed all the fluids and performed patient precision.The customer was subsequently able to run calibration and controls without issue.No further action was required by field service.A 13 month complaint history review and service history review for similar complaints was performed for the serial number (b)(4) from (b)(6) 2017 through aware date (b)(6) 2018.There were no similar complaints identified during the search period.The g8 service manual, section 6- sampling needle replacement states the following: the 6.4.1 sampling needle replacement.The sampling needle replacement requires access to the inside of the analyzer.Be sure to wear protective clothing (goggles, gloves, mask, etc.) and take sufficient care to prevent infection during handling.Follow the procedure on page 6-5, "6.3.3 needle cover removal", and remove the needle cover.You will see the sampling needle unit back in the middle.Grasp the upper part of the sampling needle unit by hand and slowly pull the unit forward as much as possible.Since a small volume of reagent will spill during replacement, place a lab wipe under the sampling needle tip.By hand, loosen and remove the joint connected to the 3-way block.Remove the screws on the upper section of the sampling needle.Be careful not to drop the screws or the holding plate inside the machine during this operation.Remove the screws holding the stay, through which the tubing passes.Slowly lift up the sampling needle to remove.Insert the new sampling needle and secure the upper plate with the screws.When you do so, make sure that the needle tip hole is oriented forward.Pass the tubing through the stay, secure with the screw, and securely connect the joint to the 3-way block.Move the sampling needle unit back and forth and confirm that the tubing does not catch anything.If necessary, loosen the screws, turn the sampling needle, and change the stay direction to prevent the tubing from catching anything.Push the sampling needle unit back and attach the needle cover.Keep the following points in mind during replacement: be sure to wear protective clothing (goggles, gloves, mask, etc.) and take sufficient care to prevent infection during sampling needle replacement.Take care not to touch the end of the sampling needle.If the needle is bent immediately after replacement, check that the primary tubes match the sample rack or sample rack adapter.If the needle placement is clearly off center of the primary tube, it must be adjusted.Change the "y-samp" parameter.Refer to page 4-16, chapter 4.Dispose of the used sampling needle as infectious waste according to the procedures at your facility.Total area: the g8 variant analysis mode operator's manual under chapter 1, introduction and applications, states the following: the 1.8 limitations of the procedure.Total area.Dilution studies demonstrate that the assay is linear from a total area of 500 to 4000.However, the optimum total area is 700 to 3000.Interpretation of results: results will not be reported if the total area (ta) is <500 which can be seen in severe anemia.Results will not be reported if the ta is >4000 which can be seen in polycythemia.(see "abnormal red cell survival in previous section).The optimal goal for total area is between 700-3000.However a ta in the range of 500-4000 is acceptable and reportable for whole blood specimens.The most probable cause of the reported event was due to fault/ failure of the sampling needle assembly.
|