Model Number 03.816.012
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Code Available (3191)
Event Date 03/27/2018
Event Type  Injury  
Manufacturer Narrative
Patient weight not available for reporting.Device is an instrument and is not implanted/explanted.Date returned to manufacturer.Initial reporter is synthes sales consultant.(b)(4).Dhr review was completed.Part: 03.816.012.Lot: 9345834.Manufacturing site: (b)(4).Release to warehouse date: 01.June 2015.The device history record shows this lot of 24 pieces was processed through the normal manufacturing and inspection operations with no rework or nonconformities noted.This lot met all dimensional, visual and functional criteria at the time of release with no issues documented during the manufacturing process.The manufacturing documents were reviewed and no complaint related issues were found., : customer quality conducted an investigation of the returned device.Investigation flow: device interaction/functional visual inspection showed that the instrument for disc anchor blade, 03.816.012, was assembled with the disc anchor, 03.816.330.Upon further inspection it was observed that the device is over inserted past the proximal edge of the blade in the disc anchor.It is probable that this condition has led to deformation of the dove tail feature on the mating surfaces of the two devices resulting in the stuck condition.The instrument for disc anchor blade, 03.816.012, is intended to assemble up to the proximal edge of the blade in the disc anchor after which the knob on the proximal end of the device is rotated to advance the blade.However, witness marks consistent with impact were observed on the concomitant knob, 03.816.014, which is consistent with forceful advancement of the device potentially leading to the over insertion functional inspection could not be performed as the instrument for disc anchor blade, 03.816.012, and the disc anchor, 03.816.330, could not be disassembled.The complaint condition is confirmed, agrees with the complaint condition of ¿device interaction (2+ devices): unable to disassemble : rm¿, and could be replicated as the devices could not be disassembled.During document/specification review the following drawings, reflecting the current and manufactured revision, were reviewed.Disc anchor instrument: the design history was found to not impact the complaint condition.During the investigation no product design issues or discrepancies were observed that may have contributed to the complaint condition.Dimensional inspection of relevant features could not be conducted due to the inability to disassemble the devices.The insight lateral access system technique guide provides instruction on use of the device and assembly/disassembly of the construct.Based on the received condition, the probable cause of the issue was determined to be the over insertion of the instrument into the disc anchor blade.The system risk document was found to adequately address the complaint condition.In conclusion, the complaint condition is confirmed as the instrument for disc anchor blade, 03.816.012, and the disc anchor, 03.816.330, could not be disassembled.No design or manufacturing defect or deficiency was observed during the investigation.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that during an unknown surgery on (b)(6) 2018, the disc anchor blade was stuck on the anchor blade of an insight lateral access retractor system.The surgeon attempted to use a back up insight lateral access retractor system and experienced the same event where the disc anchor blade was stuck to the anchor blade of the insight lateral access retractor system.Since the two (2) retractors failed, the surgeon went with a competitor¿s retractor.The procedure was completed using the competitor¿s retractor to access the disc space, but the depuy synthes implants were still used as an inter-body.It was unknown if there was a surgical delay.As reported, the patient has been doing well.Concomitant devices reported: knob for disc anchor instrument (03.816.014, lot l172936, quantity 1), third blade holder (03.816.004, lot 9275792, quantity 1), third blade holder (03.816.004, lot 9345894, quantity 1) this report is for one (1) instrument for disc anchor this is report 6 of 6 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSTRUMENT FOR DISC ANCHOR
Type of Device
RETRACTOR
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
Manufacturer (Section G)
SYNTHES HAEGENDORF GMBH-CN
im bifang 6
haegendorf 4614
SZ   4614
Manufacturer Contact
michael cote
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key7545807
MDR Text Key109313542
Report Number2939274-2018-52315
Device Sequence Number1
Product Code GAD
UDI-Device Identifier10705034730684
UDI-Public(01)10705034730684
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 03/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/25/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number03.816.012
Device Catalogue Number03.816.012
Device Lot Number9345834
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/23/2018
Date Manufacturer Received05/23/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/01/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age54 YR
-
-