LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 16-02-85
Device Problems Failure to Calibrate (2440); Device Displays Incorrect Message (2591)
Patient Problem No Patient Involvement (2645)
Event Date 04/27/2018
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).The reported issue was identified by the livanova technician during routine maintenance.The technician was able to trace the reported issue to a faulty temperature probe and processor board.Both parts were replaced to resolve the reported issue.Subsequent functional verification testing was completed without further issues and the unit was returned to service.The replaced parts have been requested for return to livanova (b)(4) for further investigation.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that a heater-cooler system 3t could not be calibrated and displayed an error message.This issue was identified by a livanova field service representative during routine maintenance.There was no patient involvement.
 
Manufacturer Narrative
The board and the sensor were returned for further investigation.During the evaluation the reported issue could not be confirmed or reproduced.Both parts worked according the specification.As the issue could not be reproduced or confirmed, a root cause was not identified.
 
Event Description
See initial.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich
MDR Report Key7546123
MDR Text Key109314919
Report Number9611109-2018-00991
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
PMA/PMN Number
K052601
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Remedial Action Other
Type of Report Initial,Followup
Report Date 01/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/25/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Medical Equipment Company Technician/Representative
Device Model Number16-02-85
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/07/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-