Model Number 16-02-85 |
Device Problems
Failure to Calibrate (2440); Device Displays Incorrect Message (2591)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 04/27/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
There was no patient involvement.Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).The reported issue was identified by the livanova technician during routine maintenance.The technician was able to trace the reported issue to a faulty temperature probe and processor board.Both parts were replaced to resolve the reported issue.Subsequent functional verification testing was completed without further issues and the unit was returned to service.The replaced parts have been requested for return to livanova (b)(4) for further investigation.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova (b)(4) received a report that a heater-cooler system 3t could not be calibrated and displayed an error message.This issue was identified by a livanova field service representative during routine maintenance.There was no patient involvement.
|
|
Manufacturer Narrative
|
The board and the sensor were returned for further investigation.During the evaluation the reported issue could not be confirmed or reproduced.Both parts worked according the specification.As the issue could not be reproduced or confirmed, a root cause was not identified.
|
|
Event Description
|
See initial.
|
|