MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II PUMP, INFUSION, IMPLANTED, PROGRAMMABLE Back to Search Results
Model Number 8637-20
Device Problems Electromagnetic Interference (1194); Mechanical Jam (2983)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/09/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a healthcare professional (hcp) regarding a patient receiving bupivacaine 15mg/ml for a total dose of 3. 913 mg/day and morphine 30 mg/ml for a total dose of 7. 827mg/day via an implantable pump for non-malignant pain. It was reported a motor stall was seen at initial interrogation. The patient recently had an magnetic resonance imaging (mri). It was noted a motor stall recovery did not occur. It was noted that the patient had an mri on (b)(6) 2018 and on (b)(6) 2018. The logs shoed a motor stall and recovery on (b)(6) 2018, but no recovery after the stall on (b)(6) 2018. It was noted it had not been two hours since the patient exited the mri field. Telemetry state and re-interrogating the pump with logs were reviewed. It was noted that no interrogations resulted in a recovery. It was advised to continue to check logs for motor stall recovery with expectation that it will show in logs with today's ((b)(6) 2018) date. No symptoms were reported. The event date was (b)(6) 2018. No further complications were reported/anticipated.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSYNCHROMED II
Type of DevicePUMP, INFUSION, IMPLANTED, PROGRAMMABLE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key7546563
MDR Text Key109381886
Report Number3004209178-2018-11956
Device Sequence Number1
Product Code LKK
UDI-Device Identifier00643169530119
UDI-Public00643169530119
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P860004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date 10/05/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received05/25/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date07/14/2019
Device Model Number8637-20
Device Catalogue Number8637-20
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received08/08/2018
Was Device Evaluated by Manufacturer? No Answer Provided
Date Device Manufactured01/22/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-0591-2009

-
-