Model Number FUSION
Device Problem Erratic or Intermittent Display (1182)
Patient Problem No Patient Involvement (2645)
Event Date 05/03/2018
Event Type  malfunction  
Manufacturer Narrative
A medtronic representative went to the site to test the equipment.The representative replaced the monitor and cables.The navigation system then passed the system checkout and was found to be fully functional.The suspect monitor was returned for evaluation.Analysis was unable to confirm the reported complaint.There was no fault found.
 
Event Description
Medtronic received information regarding a navigation device being used outside a procedure.It was reported that the monitor display goes in and out when the system arm was moved around.The site had to wiggle the display cable to bring the image back up on the screen.There was no patient present when this issue was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FUSION NAVIGATION SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
sarah olhoft
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
MDR Report Key7559826
MDR Text Key109741862
Report Number1723170-2018-02389
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00613994380005
UDI-Public00613994380005
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001284
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Nurse
Type of Report Initial
Report Date 06/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/01/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberFUSION
Device Catalogue Number9733560XOM
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/29/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/03/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/19/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-