• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WILLIAM COOK EUROPE COOK CELECT® PLATINUM NAVALIGN JUGULAR VENA CAVA FILTER SET; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WILLIAM COOK EUROPE COOK CELECT® PLATINUM NAVALIGN JUGULAR VENA CAVA FILTER SET; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR Back to Search Results
Model Number G34309
Device Problem Positioning Failure (1158)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/16/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).(b)(4).Catalog#: igtcfs-65-1-jug-celect-pt (b)(6).Investigation is still in progress.
 
Event Description
Description of event according to initial reporter: "the filter failed to deploy.A new one was opened and used to complete the procedure".Patient outcome: the patient did not require any additional procedures due to this occurrence.No adverse effects were reported on the patient due to this occurrence.
 
Event Description
Description of event according to initial reporter: "the filter failed to deploy.A new one was opened and used to complete the procedure".Patient outcome: the patient did not require any additional procedures due to this occurrence.No adverse effects were reported on the patient due to this occurrence.
 
Manufacturer Narrative
Exemption number e2016032.William cook europe aps (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi) (importer).Manufacturer ref# pr228002.G1) name and address for importer site: cook medical incorporated (cmi) 400 daniels way bloomington, in 47404 registration no.: 3005580113 h6) ec method code: 4114 - device not returned ec conclusions code: 4315 - cause not established.Summary of investigational findings: assume "the filter failed to deploy" means the filter would not expand.No product returned and without the actual complaint device, it would be inappropriate to speculate at what may or may not have prevented the filter from expanding.However, it is previously seen that the filter legs may be somehow obstructed from fully expanding due to e.G.Ivc anatomical conditions, clots or if the filter is not placed in ivc.The physician succesfully removed the filter and the patient was implanted with new filter.No evidence to suggest that this device was not manufactured according to specifications cook medical will continue to monitor for similar events.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COOK CELECT® PLATINUM NAVALIGN JUGULAR VENA CAVA FILTER SET
Type of Device
DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
WILLIAM COOK EUROPE
sandet 6
bjaeverskov 4632
DA  4632
MDR Report Key7559878
MDR Text Key109905630
Report Number3002808486-2018-00653
Device Sequence Number1
Product Code DTK
UDI-Device Identifier10827002343099
UDI-Public(01)10827002343099(17)201120(10)E3656483
Combination Product (y/n)N
PMA/PMN Number
K121629
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 08/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/01/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/20/2020
Device Model NumberG34309
Device Catalogue NumberIGTCFS-65-1-JUG-CELECT-PT
Device Lot NumberE3656483
Was Device Available for Evaluation? No
Distributor Facility Aware Date05/17/2018
Device Age6 MO
Date Manufacturer Received07/24/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-