Device Problem Component Falling (1105)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/04/2018
Event Type  malfunction  
Manufacturer Narrative
The lighting system was installed in (b)(6) 2016 and is not under a steris service contract for preventive maintenance.Maintenance activities are provided by the facility's biomed department.Immediately following the event, the facility's biomed department replaced the lighting system covers with inventory available in stock, tested the unit, and returned it to service.A steris service technician offered to inspect the lighting system and lighting covers following the event; however, the offer was declined.The customer stated that both pieces of the lighting system cover were damaged and had broken at the screw connecting point where the screw is installed to secure the covers to the lighting suspension arm.Based on the description of the event, it appears the detachment of the lighting system covers is attributed to impact damage from the lighting system contacting other equipment in the or environment.The harmony air g5 surgical lighting system operator manual includes a list of safety precautions which include: "caution - possible equipment damage: do not bump lightheads into walls or other equipment".Following the event, the facility's steris account manager counseled the facility on the importance of adhering to the safety precautions outlined in the unit's operator manual and provided the facility with an extra copy of the applicable warnings.No further issues have been reported.
 
Event Description
The user facility reported the lighting covers on their harmony air g5 surgical lighting system fell during patient procedure setup.The covers did not contact the patient or user facility staff.No injury, procedure delay, or cancellation occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HARMONY AIR G5 HD CAMERA READY PACK
Type of Device
SURGICAL LIGHTING SYSTEM
Manufacturer (Section D)
STERIS CORPORATION - MONTGOMERY
2720 gunter park drive east
montgomery AL 36109
Manufacturer (Section G)
STERIS CORPORATION - MONTGOMERY
2720 gunter park drive east
montgomery AL 36109
Manufacturer Contact
daniel davy
5960 heisley road
mentor, OH 44060
4403927453
MDR Report Key7561014
MDR Text Key109907343
Report Number1043572-2018-00041
Device Sequence Number1
Product Code FSY
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 06/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/01/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received05/04/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-