MEDTRONIC MED REL MEDTRONIC PUERTO RICO RESTORE ULTRA; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF Back to Search Results
Model Number 37712
Device Problems Improper or Incorrect Procedure or Method (2017); Low Battery (2584); Charging Problem (2892)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/01/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a manufacturer¿s representative (rep) regarding a patient who was implanted with an implantable neurostimulator (ins) for multiple back operations.The rep reported that the healthcare provider (hcp) called the rep to set up an appointment to check the battery because the patient couldn¿t charge her ins.The rep reported that the battery was 8 years old and the patient¿s history of previous overdischarges was unknown.The rep reported that they were seeing the patient (b)(6) 2018.The issue wasn¿t resolved.No further complications were reported.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information was received from the manufacturer's representative.It was reported that cause of the ins not charging, was due to patient not having charged for several months.The rep reported they verified the overdischarge and advised the patient that due to the age of the battery and history of previous od's it would probably be necessary to replace the battery.Patient is going to see a surgeon for a consult to get the battery replaced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESTORE ULTRA
Type of Device
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF
Manufacturer (Section D)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key7561358
MDR Text Key109898302
Report Number3004209178-2018-12384
Device Sequence Number1
Product Code LGW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P840001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/01/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/14/2011
Device Model Number37712
Device Catalogue Number37712
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received06/21/2018
Date Device Manufactured03/15/2010
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age72 YR
-
-