Intuitive surgical, inc.(isi) received the instrument involved with this complaint and completed the device evaluation.Failure analysis confirmed the customer reported failure.The instrument was found to have a broken pitch cable at the distal clevis hub.Crimp was missing.Isi has conducted a device history record (dhr) review for this device and did not find any non-conformances that were related to this reported event.The customer reported complaint does not itself constitute an mdr reportable event; however, the broken pitch cable found during failure analysis could cause or contribute to an adverse event if the malfunction were to recur.
|
It was reported that during a da vinci-assisted surgical procedure, the grasping retractor instrument had a "sprung wire" at the tip.There was no report of fragment(s) falling inside the patient, patient harm, adverse outcome or injury.
|