Model Number S7 |
Device Problem
No Device Output (1435)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 05/09/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
No patient information provided as no patient was involved in this concern.The suspect part 24v battery was returned to the manufacturer for evaluation.However, results are not available at this time.
|
|
Event Description
|
Medtronic received information that, while outside of a procedure, the navigation system would power down after being unplugged from the outlet.It was noted the battery was not able to hold a charge.There was no patient present when this issue was identified.
|
|
Manufacturer Narrative
|
The analysis on the battery was completed by medtronic personnel.Testing found that the battery would not pass load testing.The hardware investigation found that the reported event was related to a hardware issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.
|
|
Manufacturer Narrative
|
A medtronic representative went to the site to test the equipment.Testing revealed that after replacing the battery the system functioned as intended.The system then passed the system checkout and was found to be fully functional.If information is provided in the future, a supplemental report will be issued.
|
|