• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - GALWAY SYNERGY¿; CORONARY DRUG-ELUTING STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - GALWAY SYNERGY¿; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number H7493926232220
Device Problem Bent (1059)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/15/2018
Event Type  malfunction  
Manufacturer Narrative
Age at time of event: 18 years or older.Device is a combination product.(b)(4).Device evaluated by mfr: the device was not returned for analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is operational context as the product meets the design & manufacture specification but due to anatomical/procedural factors encountered during the procedure, performance was limited.(b)(4).
 
Event Description
It was reported that stent damage occurred.Vascular access was obtained via femoral artery.The 90% stenosed, 12-32mm x 2.25-2.5mm, eccentric, de novo target lesion with a bend between 45 and 90 degrees was located in the mildly tortuous and mildly calcified coronary vessel.A 2.25 x 32 synergy¿ stent was advanced but failed to cross the lesion.The device was removed and it was noted that the stent was deformed.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY¿
Type of Device
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
Manufacturer (Section G)
BOSTON SCIENTIFIC - GALWAY
Manufacturer Contact
sonali arangil
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key7570612
MDR Text Key110123502
Report Number2134265-2018-04932
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
SIMILAR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial
Report Date 05/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/05/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/23/2019
Device Model NumberH7493926232220
Device Catalogue Number39262-3222
Device Lot Number0020997415
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/15/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/27/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-