• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - MOUNTAIN HOME HOMECHOICE AUTOMATED PD SET WITH CASSETTE; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - MOUNTAIN HOME HOMECHOICE AUTOMATED PD SET WITH CASSETTE; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number L5C4531
Device Problem Cut In Material (2454)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/10/2018
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that there were slits along the patient line of a homechoice cassette that were visible through the overpouches.This was discovered before use of the device.The home patient also described "gouges in the tubing".There was no report of leaks.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
There were five (5) actual samples received for evaluation, four (4) in opened pouches and one (1) in an un-opened pouch.Visual inspection with naked eye and magnification identified surface marks on the patient line next to patient connector on five (5) units and several small circle identations on the tubing of one (1) unit.Functional testing including leak testing, clear passage testing and clamp function testing were performed with no issues noted.The reported issue was verified.The cause was determined to be manufacturing related.The surface marks on the patient line tubing were caused by the coiling head transfer gripper during the automated assembly system.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE - MOUNTAIN HOME
mountain home AR
Manufacturer (Section G)
BAXTER HEALTHCARE - MOUNTAIN HOME
1900 n highway 201
mountain home AR 72653
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key7571316
MDR Text Key110367050
Report Number1416980-2018-03424
Device Sequence Number1
Product Code FKX
UDI-Device Identifier00085412090078
UDI-Public(01)00085412090078
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102936
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/05/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date03/14/2023
Device Catalogue NumberL5C4531
Device Lot NumberH18C14039
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/24/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received07/13/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Age56 YR
-
-