• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CONCORD MANUFACTURING 2008K2 HEMODIALYSIS SYS. OLC/DIASAFE PLS DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONCORD MANUFACTURING 2008K2 HEMODIALYSIS SYS. OLC/DIASAFE PLS DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 190610
Device Problems Thermal Decomposition of Device (1071); Corroded (1131)
Patient Problem No Patient Involvement (2645)
Event Date 05/15/2018
Event Type  malfunction  
Manufacturer Narrative
Plant investigation: although it was stated that the complaint device was available to be returned to the manufacturer for physical evaluation, to date none has been received. The reported complaint was not confirmed as the complaint device was not returned to the manufacturer for evaluation. As such, a definitive conclusion regarding the complaint incident cannot be reached without a physical examination of the complaint device. A review of the device manufacturing records was conducted by the manufacturer. There were no deviations or non-conformances during the manufacturing process. In addition, a device history record (dhr) review was performed and verified that the results of the in-progress and final quality control (qc) testing met all requirements. The investigation into the cause of the reported problem was not able to be confirmed. A definitive conclusion regarding the complaint incident cannot be reached without a physical examination of the complaint device. Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
A biomedical technician (biomed) at a user facility reported that while troubleshooting the 2008k2 hemodialysis machine for conductivity issues, it was noted that the wiring harness was corroded due to heat damage. The biomed stated that the insulation on the temperature sensors was melted due to heat damage and eventually became brittle and was crumbling off. The biomed stated that the temperature control thermistor insulations were melted and burnt. The biomed stated that it is unsure if a patient was on the machine or not at the time of the conductivity issues; however confirmed that there was no adverse events, medical intervention or injuries reported. The biomed replaced the temperature sensors. The machine has been returned to service at the user facility without recurrence of the reported issue. The complaint device is available to be returned to the manufacturer for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name2008K2 HEMODIALYSIS SYS. OLC/DIASAFE PLS
Type of DeviceDIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
CONCORD MANUFACTURING
4040 nelson avenue
concord CA 94520
Manufacturer (Section G)
CONCORD MANUFACTURING
4040 nelson avenue
concord CA 94520
Manufacturer Contact
thomas c. johnson
920 winter st.
waltham, MA 02451
7816999499
MDR Report Key7572878
MDR Text Key110181302
Report Number2937457-2018-01585
Device Sequence Number1
Product Code KDI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K994267
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation
Type of Report Initial
Report Date 06/06/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received06/06/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue Number190610
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Device AgeMO
Event Location No Information
Date Manufacturer Received05/24/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/27/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-