Model Number DGN-538-5
Device Problem Detachment Of Device Component (1104)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/15/2018
Event Type  Injury  
Manufacturer Narrative
Patient's exact age is unknown; however it was reported that the patient was over the age of 18.The complainant was unable to provide the suspect device lot number; therefore, the lot expiration and device manufacture dates are unknown.However, the complainant reported that the device was not expired.(b)(4) relates to the reported event of net detached.(b)(4).According to the complainant, the suspect device has been disposed and is not available for return.If any further relevant information is received, a supplemental mdr will be filed.
 
Event Description
This report pertains to the first of three devices used during the same procedure.Refer to manufacturer report # 3005099803-2018-01823 and 3005099803-2018-01824 for the other associated device information.It was reported to boston scientific corporation that three rescuenet retrieval devices were used in the esophagus during a food bolus retrieval procedure performed on (b)(6) 2018.According to the complainant, during the procedure, all three nets completely detached from the loop and the wire inside the catheter broke.The detached component was retrieved successfully with a different device and the procedure was completed with another rescuenet.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESCUENET¿ RETRIEVAL NET
Type of Device
RETRIEVAL NET
Manufacturer (Section D)
BOSTON SCIENTIFIC - COSTA RICA (COYOL)
2546 first street
propark free zone
alajuela
CS 
Manufacturer Contact
nancy cutino
100 boston scientific way
marlborough, MA 01752
5086834000
MDR Report Key7574985
MDR Text Key110320971
Report Number3005099803-2018-01822
Device Sequence Number1
Product Code FDI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K152580 
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 05/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/06/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberDGN-538-5
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/15/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-