• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER CONSTELLATION VISION SYSTEM; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER CONSTELLATION VISION SYSTEM; UNIT, PHACOFRAGMENTATION Back to Search Results
Model Number TABLETOP
Device Problems Air Leak (1008); Device Displays Incorrect Message (2591); Aspiration Issue (2883)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/03/2018
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
 
Event Description
A nurse reported that during a surgical procedure the system displayed a system message to release the foot pedal.Bubbles appeared in the suction tubing and there was no aspiration.The case was completed without harm to the patient.Additional information has been requested but not received.
 
Manufacturer Narrative
The company service representative examined the system but was unable to replicate the problem reported.Unrelated to the reported event, a software upgrade was performed, and a fan kit was installed.The system was then tested and met all product specifications.The system was manufactured on november 25, 2015.Based on qa assessment, the product met specifications at the time of release.The customer did not retain a cassette pak sample for this complaint report; visual inspection or functional testing could not be conducted.The lot number complaint history and manufacturing record were not reviewed since the lot number was not provided.The root cause of the reported events cannot be determined conclusively.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
bryan blake
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8176152230
MDR Report Key7579670
MDR Text Key110715658
Report Number2028159-2018-01173
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K101285
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 08/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/07/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberTABLETOP
Device Catalogue Number8065751150
Other Device ID Number380657511501
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/01/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/25/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-