Catalog Number RF400F
Device Problem Extrusion (2934)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  Malfunction  
Manufacturer Narrative

Manufacturing review: as the lot number for the device was not provided, a manufacturing review could not be performed. Medical records review: the patient with history of pulmonary embolus had a prophylactic inferior vena cava (ivc) filter deployed prior to left knee replacement surgery. The patient tolerated the procedure well and was discharged home the same day. Investigation summary: the device was not returned for evaluation. Images were not provided for review. Medical records were provided and reviewed. There was no specific deficiency alleged in the provided medical records. Therefore, the investigation is inconclusive for perforation as no objective evidence has been provided to confirm any alleged deficiency with the filter. Based upon the available information, the definitive root cause is unknown. Labeling review: a review of product labeling documents (e. G. Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, and unit label) showed that the product labeling is adequate. The information provided by bard represents all of the known information at this time. Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.

 
Event Description

It was reported through the litigation process that a vena cava filter was placed in a patient in conjunction with or before orthopedic procedure and the patient had a history of deep vein thrombosis. At some time post filter deployment, it was alleged that filter struts perforated into organs. The device was removed percutaneously. The status of the patient is unknown.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameG2X FILTER SYSTEM - FEMORAL
Type of DeviceVENA CAVA FILTER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
C.R. BARD, INC. (GFO)
289 bay road
queensbury NY 12804
Manufacturer Contact
judith ludwig
1625 w 3rd st.
tempe, AZ 85281
4803032689
MDR Report Key7580766
MDR Text Key110455659
Report Number2020394-2018-00820
Device Sequence Number1
Product Code DTK
Combination Product (Y/N)N
Reporter Country CodeUS
PMA/PMN NumberK082305
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type CONSUMER,HEALTH PROFESSIONAL
Reporter Occupation ATTORNEY
Type of Report Initial
Report Date 06/07/2018
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received06/07/2018
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device Catalogue NumberRF400F
Device LOT NumberUNKNOWN
Was Device Available For Evaluation? No
Is The Reporter A Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received05/14/2018
Was Device Evaluated By Manufacturer? Device Not Returned To Manufacturer
Is The Device Single Use? Yes
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-