A medtronic representative went to the site to test the equipment.
The hardware, software, and instruments passed the system checkout.
The system was found to be fully functional.
The computer was returned to the manufacturer for analysis.
Analysis found that system passed sys test and all refurbishment requirements.
Analysis found that there was no fault with the system.
|
Medtronic received information regarding a navigation device being used for a functional endoscopic sinus surgery (fess) procedure.
It was reported that intraoperatively during the navigation task, the software jumped back to the select patient screen again.
This occurred approximately 1.
5 hours into the case.
The frazier suction was inside the patient¿s nose when the system jumped back.
It was confirmed with the manufacturer representative that the site had reinstalled the software since the last case where a similar issue occurred.
This issue resulted in a delay of surgery of less than 1 hour, and there was no impact on patient outcome.
|