ZIMMER BIOMET, INC. ARCOM XL 44-36 STD HMRL BRNG; PROSTHESIS, SHOULDER Back to Search Results
Model Number N/A
Device Problems Difficult to Insert (1316); Difficult To Position (1467)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/09/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2018-03777.
 
Event Description
It was reported that during a right shoulder arthroplasty, the bearing did not engage the humeral tray correctly and locking ring was found jammed upon inspection.Same size implants were used to replace intraoperatively, and no harm to the patient was indicated.Attempts to obtain additional information have been made; however, no more is available at this time.
 
Manufacturer Narrative
Reported event is considered not confirmed as the returned device met all specifications; the returned the bearing was properly assembled to the tray with no noted damage and no gaps identified.Dhr was reviewed and no discrepancies were found.Root cause could not be determined as there was no problem identified with the returned product.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCOM XL 44-36 STD HMRL BRNG
Type of Device
PROSTHESIS, SHOULDER
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key7581861
MDR Text Key110463174
Report Number0001825034-2018-03819
Device Sequence Number1
Product Code KWS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK080642
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/08/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date02/27/2023
Device Model NumberN/A
Device Catalogue NumberXL-115363
Device Lot Number875000
Other Device ID Number(01) 0 0880304 47542 7
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/31/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/10/2018
Was Device Evaluated by Manufacturer? No
Date Device Manufactured02/27/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
115370 COMP RVS TRAY CO 44MM 350660
-
-