Model Number FUSION |
Device Problems
Device Stops Intermittently (1599); Device Operates Differently Than Expected (2913)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 05/11/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
No patient information provided as no patient was involved in this concern.
A medtronic representative went to the site to test the equipment.
It was unable to duplicate the issue; however, the system failed the hardware checkout but passed the software and instrument checkout.
The emitter was replaced, and system performed as intended.
The imaging system then passed the system checkout and was found to be fully functional.
No parts have been returned to the manufacturer for evaluation.
|
|
Event Description
|
Medtronic received information regarding a navigation device being used outside of procedure.
It was reported that when trying to track the instrument tracker it will flicker in and out while moving it slowly a fists distance from the emitter.
When the patient tracker moves the same thing occurs.
|
|
Manufacturer Narrative
|
The field generator was returned to the manufacturer for analysis.
The unit was found to be fully functional with no problem found.
The reported event could not be duplicated by medtronic personnel.
|
|