U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CONCORD MANUFACTURING 2008T HEMODIALYSIS SYS., WITH CDX DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONCORD MANUFACTURING 2008T HEMODIALYSIS SYS., WITH CDX DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 190713
Device Problems Charred (1086); Melted (1385)
Patient Problem No Patient Involvement (2645)
Event Date 05/10/2018
Event Type  malfunction  
Manufacturer Narrative
The plant investigation is in process. A supplemental mdr will be submitted upon completion of this activity.
 
Event Description
A biomedical technician (biomed) reported that the 2008t hemodialysis machine spade connector had sustained heat damage. The biomed visibly observed the damaged component and stated that it was a light brown color. The biomed replaced the rocker switch and wires to resolve the issue. The machine has been returned to service at the user facility. The biomed confirmed that there was no patient involvement associated with the reported event. The reported device complaint is available to be returned to the manufacturer for physical evaluation.
 
Manufacturer Narrative
Plant investigation: no parts were returned to the manufacturer for physical evaluation. Additionally, no on-site evaluation was performed by a fresenius regional equipment specialist (res). A records review was performed on the reported serial number. An investigation of the device manufacturing records was conducted by the manufacturer. There were no non-conformances, or any associated rework identified during the manufacturing process which could be related to the reported event. In addition, the device history record (dhr) review confirmed the results of the in-progress and final quality control (qc) testing met all requirements. The investigation into the cause of the reported problem was not able to be confirmed. A definitive conclusion regarding the complaint incident cannot be reached without a physical examination of the complaint device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name2008T HEMODIALYSIS SYS., WITH CDX
Type of DeviceDIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
CONCORD MANUFACTURING
4040 nelson avenue
concord CA 94520
MDR Report Key7583147
MDR Text Key110581747
Report Number2937457-2018-01660
Device Sequence Number1
Product Code KDI
Combination Product (y/n)N
PMA/PMN Number
K093902
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Type of Report Initial,Followup
Report Date 06/29/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received06/08/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue Number190713
Was Device Available for Evaluation? No
Was the Report Sent to FDA?
Device AgeMO
Event Location No Information
Date Manufacturer Received06/26/2018
Is This a Reprocessed and Reused Single-Use Device? No

-
-