Model Number 20
Device Problems Device Inoperable (1663); Connection Problem (2900)
Patient Problem No Patient Involvement (2645)
Event Date 05/01/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Physio-control continues to investigate the reported issue and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
 
Event Description
The customer contacted physio-control to report that their device had a service indicator present and would not complete the initial boot-up cycle in order to power on.As a result, defibrillation would not be possible if it were necessary.There was no patient use associated with the reported event.
 
Manufacturer Narrative
(b)(4).Physio-control evaluated the customer's device but was unable to duplicate the reported failure to complete the initial boot-up cycle in order to power on.Physio-control did, however, observe that the device was unable to detect that a test load was connected via a therapy cable.As a result, the device would not be able to defibrillate if it were necessary.Physio then replaced the system controller pcb assembly to resolve the observed issue and completed other, unrelated, repairs.After observing proper device operation through functional and performance testing the unit was returned to the customer for use.Physio-control further evaluated the removed sc pcb assembly and confirmed that it had caused the observed issue.Physio found that the pcb does not allow for proper measurement of impedance.However, further component level cause could not be determined.The cause of the original reported issue (failure to complete the boot-up cycle) could not be determined.
 
Event Description
The customer contacted physio-control to report that their device had a service indicator present and would not complete the initial boot-up cycle in order to power on.As a result, defibrillation would not be possible if it were necessary.There was no patient use associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR
Type of Device
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
megan marseglia
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key7583609
MDR Text Key110530960
Report Number0003015876-2018-00901
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K130454
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 07/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/08/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number20
Device Catalogue Number99402-000016
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/12/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/17/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/31/2003
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-