Catalog Number 8065751765
Device Problem Difficult to Open or Close (2921)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/04/2018
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
 
Event Description
A physician reported the self closing valve trocar did not completely close causing overflow of the perfusion fluid and vitreous body during a procedure.The procedure was completed by using scleral plugs.There was no harm to the patient.No additional information is expected.
 
Manufacturer Narrative
No sample has been returned for evaluation for the report of leaking trocar; therefore, the condition of the product could not be verified.A review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.A complaint history examination indicates there are sixteen additional complaints associated with the lot for the reported issue.No sample was returned and the device history record review of the lot number provided indicated product was processed and released according to the product¿s acceptance criteria, therefore the root cause for the defect experienced by the customer cannot be determined.The exact root cause for this complaint is unknown.Investigations have been completed and manufacturing process enhancements have been implemented in order to improve the performance of the valves.Complaints are reviewed and monitored at regular intervals to evaluate effectiveness of actions taken.No additional action is required at this time.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION SURGICAL PROCEDURE PAK
Type of Device
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LTD. - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LTD. - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
bryan blake
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8176152230
MDR Report Key7584231
MDR Text Key111017348
Report Number1644019-2018-00136
Device Sequence Number1
Product Code LRO
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K880961
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/08/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2018
Device Catalogue Number8065751765
Device Lot Number1909272H
Other Device ID Number380657517657
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/28/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/20/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-