Catalog Number 405482 |
Device Problem
Incorrect Measurement (1383)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 05/29/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
|
Event Description
|
It was reported with the use of the bd¿ needle bns tuohy open shld there was an issue with incorrect needle length.It was stated ¿while performing the incoming inspection for this batch, the inspector found that the length of the bevel appeared to be much longer than noted on the graphic bevel length listed as.120 +/-.005.¿ there was no report of injury or further medical intervention.
|
|
Manufacturer Narrative
|
Investigation summary: ten samples of catalog 405482 were receive to bd juncos on 20jun18.Samples included all assembly components, such as needle and stylet (17ga x 3 1/2in tuohy) and the plastic shield.Four out of ten values were out of specification.However, per bd juncos internal procedure finish point length measurement is a reference length, that it means that finish point length is not a critical specification and do not impact the quality or function of the product if the required dimensions are within spec.A lot history review was carried out and no related non conformances were raised in association with this packaged lot concluding all inspections were performed per the applicable operations and met qc specifications.Investigation conclusion: possible root cause may be related to the manufacturing process at supplier site.
|
|
Event Description
|
It was reported with the use of the bd¿ needle bns tuohy open shld there was an issue with incorrect needle length.It was stated ¿while performing the incoming inspection for this batch, the inspector found that the length of the bevel appeared to be much longer than noted on the graphic bevel length listed as.120 +/-.005.¿ there was no report of injury or further medical intervention.
|
|