Model Number MMT-751NAL |
Device Problems
Fluid/Blood Leak (1250); Device Displays Incorrect Message (2591)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 05/15/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
|
Event Description
|
The customer reported via phone call that the insulin pump alarmed compromised force sensor system.The customer¿s blood glucose level was 123 mg/dl at the time of the incident.The customer also reported that the insulin was squirting during manual prime process.The customer was advised that the insulin pump needed to be replaced and was advised to discontinue use of the insulin pump and to revert to the backup plan.The insulin pump will be returned for analysis.
|
|
Manufacturer Narrative
|
Device passed the displacement test.Device received with compromised force sensor system alarm during the basic occlusion test due to loose or protruded drive support disk.Unable to perform prime test, excessive no delivery test and occlusion test or verify prime or fill anomaly due to compromised force sensor system alarm.
|
|