Model Number FUSION
Device Problems No Display/Image (1183); Application Program Problem (2880)
Patient Problem No Patient Involvement (2645)
Event Date 05/17/2018
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.Udi not available at the time of filing.Device manufacture date is not available at time of filing.A medtronic representative went to the site to test the equipment.Testing revealed that the dicom qr application was not installed.Application installed and exams transferred.Unable to recreate issue.
 
Event Description
Medtronic received information that the monitor on this navigation system went black when the site attempts to query for a patient.They then are required to reboot the system and after rebooting, the same behavior happens.Verified the cabling is properly seated.This black screen only occurs when they attempt to query.Unique to dicom qr application.There was no patient present when this issue was identified.
 
Manufacturer Narrative
Additional information: unique device identification (udi) and device manufacture date provided.Correction: product and related fields updated to proper value.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FUSION NAVIGATION SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
peter verhey
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
MDR Report Key7591296
MDR Text Key110743091
Report Number1723170-2018-02703
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001284
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
Report Date 07/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/12/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberFUSION
Device Catalogue Number9733560XOM
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received06/25/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/12/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-