Model Number I7
Device Problems Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138); Device Stops Intermittently (1599)
Patient Problem No Patient Involvement (2645)
Event Date 05/18/2018
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.A medtronic representative went to the site to test the equipment.Testing revealed that the system became unresponsive on multiple occasions.Replacing the computer and reinstalling the software resolved the issue.The system then passed the system checkout and was found to be fully functional.The computer was returned to the manufacturer for analysis.Analysis found that the computer booted normally however bad sectors in the hard drive were found.
 
Event Description
Medtronic received information regarding an imaging device being used outside of procedure.It was reported that the system became unresponsive.The touch functions and mouse functions would not work on the monitors as well as the rack monitor.Rebooting the issue fixed it temporarily, but the issue reoccurred.There was no patient present at the time of the issue.
 
Manufacturer Narrative
Additional information: computer was tested and bad sectors on hard drive may have cause the issue.The hardware investigation found that the reported event was related to a hardware issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION I7 INTEGRATED NAVIGATION SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
prashanth gali
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
7208902515
MDR Report Key7591633
MDR Text Key110754934
Report Number1723170-2018-02709
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00613994653482
UDI-Public00613994653482
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050438
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
Report Date 07/31/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/12/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberI7
Device Catalogue Number9734060
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/05/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received07/17/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/30/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-