MEDTRONIC NAVIGATION, INC. (LOUISVILLE) FUSION NAVIGATION SYSTEM NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Model Number FUSION
Device Problems Erratic or Intermittent Display (1182); No Display/Image (1183); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/16/2018
Event Type  malfunction  
Manufacturer Narrative
Udi and manufacture date not available for this instrument at time of filing. A medtronic representative went to the site to test the equipment. Testing revealed that the monitor goes black for a few seconds then comes back on. The monitor and cables were replaced. The system then passed the system checkout and was found to be fully functional. The fusion monitor was returned to the manufacturer for analysis. It was not able to replicate the reported failure. Analysis found that when connected to a known good system the returned monitor performed as expected without any issues. No problem found.
 
Event Description
Medtronic received information that, during a functional endoscopic sinus surgery (fess) procedure, the monitor has gone black for a few seconds then come back on. There was no reported impact on patient outcome.
 
Manufacturer Narrative
Additional information: device manufacture date provided. Correction: unique device identification (udi) updated to proper value.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameFUSION NAVIGATION SYSTEM
Type of DeviceNEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
peter verhey
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
MDR Report Key7591711
MDR Text Key110763726
Report Number1723170-2018-02707
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001284
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation
Type of Report Initial,Followup
Report Date 07/17/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received06/12/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberFUSION
Device Catalogue Number9733560XOM
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/05/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received06/26/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/16/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-