Model Number S7 |
Device Problems
Communication or Transmission Problem (2896); Device Operates Differently Than Expected (2913)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 05/18/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Unique device identification (udi) is not available.Device manufacturing date is unavailable.A medtronic representative reported that the site did not have any issues since and have completed case without issue.Part not returned for evaluation.
|
|
Event Description
|
Medtronic received information regarding a navigation device being used for a sinus endoscopy functional endoscopic sinus surgery (fess) procedure.The issue occurred intraoperatively while navigating and delayed the surgery by less than an hour.It was reported that registration was completed without any issues, but in navigate, they were having difficulty tracking.The instruments were flickering in and out.The site tried reducing metal as much as possible without resolution.The site switched emitters and the issue was resolved.The surgery was completed with navigation and there was no known impact on patient outcome.
|
|
Manufacturer Narrative
|
Additional information: unique device identification (udi) and device manufacture date provided.A medtronic representative went to the site to test the equipment.Testing revealed that the reported issue could not be replicated.The system then passed the system checkout and was found to be fully functional.
|
|