DEPUY SYNTHES PRODUCTS LLC BATTERY REAMER/DRILL; MOTOR, SURGICAL INSTRUMENT, AC-POWERED
|
Back to Search Results |
|
Catalog Number 530.605 |
Device Problems
Device Stops Intermittently (1599); Mechanical Jam (2983)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 01/01/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Device was used for treatment, not diagnosis.The actual device was returned for evaluation.Reliability engineering evaluated the device and the reported condition that the device was not working was confirmed.An assessment was performed which found that the housing of the device was cracked, the trigger was sticking, and the device was running intermittently.During service and repair it was observed that the housing was cracked, the trigger assembly was corroded, the gears and gear bow were corroded, and the contacts on the motor were corroded.The assignable root causes of these conditions were determined to be due to improper handling and improper cleaning methods, which is user error/misuse/abuse.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Event Description
|
It was reported from (b)(6) that the battery reamer / drill device was not working.During in-house engineering evaluation it was determined that the housing of the device was cracked and running intermittently, and the trigger was sticking.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was not reported, however, it was reported that the event occurred in 2018.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|
Search Alerts/Recalls
|
|
|