• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES PRODUCTS LLC BATTERY REAMER/DRILL; MOTOR, SURGICAL INSTRUMENT, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES PRODUCTS LLC BATTERY REAMER/DRILL; MOTOR, SURGICAL INSTRUMENT, AC-POWERED Back to Search Results
Catalog Number 530.605
Device Problems Device Stops Intermittently (1599); Mechanical Jam (2983)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/01/2018
Event Type  malfunction  
Manufacturer Narrative
Device was used for treatment, not diagnosis.The actual device was returned for evaluation.Reliability engineering evaluated the device and the reported condition that the device was not working was confirmed.An assessment was performed which found that the housing of the device was cracked, the trigger was sticking, and the device was running intermittently.During service and repair it was observed that the housing was cracked, the trigger assembly was corroded, the gears and gear bow were corroded, and the contacts on the motor were corroded.The assignable root causes of these conditions were determined to be due to improper handling and improper cleaning methods, which is user error/misuse/abuse.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Event Description
It was reported from (b)(6) that the battery reamer / drill device was not working.During in-house engineering evaluation it was determined that the housing of the device was cracked and running intermittently, and the trigger was sticking.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was not reported, however, it was reported that the event occurred in 2018.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BATTERY REAMER/DRILL
Type of Device
MOTOR, SURGICAL INSTRUMENT, AC-POWERED
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
SYNTHES PRODUKTIONS GMBH
hauptstrasse 24
waldenburg 4437
SZ   4437
Manufacturer Contact
michael cote
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key7592885
MDR Text Key110848933
Report Number8030965-2018-54265
Device Sequence Number1
Product Code GEY
UDI-Device Identifier7611819137103
UDI-Public(01)7611819137103(11)111206
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 03/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/12/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number530.605
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/01/2018
Date Manufacturer Received05/23/2018
Date Device Manufactured12/06/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-