Catalog Number 9735665 |
Device Problems
Device Stops Intermittently (1599); Communication or Transmission Problem (2896); Device Operates Differently Than Expected (2913)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 05/21/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
A medtronic representative went to the site to test the equipment.
The hardware, software, and instruments passed the system checkout.
The system was found to be fully functional.
No parts have been returned to the manufacturer for evaluation.
|
|
Event Description
|
Medtronic received information regarding a navigation device being used for an electrode and probe placement procedure.
It was reported that images were unmerged from the original imaging system.
Site attempted to merge into another imaging system but localizer failed.
Patient was re-positioned to truer ap-localizer passed.
The imaging system was merged with the mri, but entire patient file was wiped from the system.
Patient files were reloaded from planning station.
The imaging system pushed to the navigation system.
Images merged and localizer passed, however, when entering full screen all plans were labeled left even though 7 of the 9 images should have been right.
The site completed the procedure without further complication.
There was no impact to patient.
There was a reported delay to the procedure of less than 1 hour due to this issue.
|
|
Manufacturer Narrative
|
A software analysis was initiated to determine the probable cause of the issue through archive/image analysis.
Analysis found that the reported event was related to a software issue.
This issue was documented in a medtronic navigation software anomaly tracking database.
|
|