Model Number 20
Device Problem Device Inoperable (1663)
Patient Problem No Patient Involvement (2645)
Event Date 05/23/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Physio-control evaluated the customer¿s device and verified the reported issue.Physio-control replaced the user interface pcb assembly.After other unrelated repairs were completed, proper device operation was observed through functional and performance testing and the device was returned to the customer for use.Physio-control further evaluated the removed user interface pcb assembly.It was determined that the memory of integrated circuit chip, designator u2 had become corrupt and caused the reported issue.
 
Event Description
The customer contacted physio-control to report a non-critical issue with their device.Upon evaluation of the customer¿s device, physio-control observed that the device intermittently was unable to complete a boot cycle and displayed a white screen.In this state the device would be inoperable and defibrillation therapy would not be available if needed.There was no patient use associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR
Type of Device
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
megan marseglia
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key7598880
MDR Text Key111008298
Report Number0003015876-2018-00932
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K130454
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 06/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/13/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number20
Device Catalogue Number99402-000016
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/30/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/08/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/29/2004
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-