MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. G2 FILTER SYSTEM; VENA CAVA FILTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. G2 FILTER SYSTEM; VENA CAVA FILTER Back to Search Results
Catalog Number UNKNOWN G2
Device Problems Extrusion (2934); Patient Device Interaction Problem (4001)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/25/2017
Event Type  malfunction  
Manufacturer Narrative
Manufacturing review: a complete manufacturing review could not be conducted for the investigation as the lot number is unknown.Investigation summary: the device was not returned for evaluation.Images and medical records were not provided for review.Therefore, the investigation is inconclusive for perforation of the ivc as no objective evidence has been provided to confirm any alleged deficiency with the filter.Based upon the available information, the definitive root cause is unknown.Labeling review: a review of product labeling documents (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, and unit label) showed that the product labeling is adequate.
 
Event Description
It was reported through the litigation process that a vena cava filter was placed in a patient after being diagnosed with deep vein thrombosis/pulmonary embolism.At some time post filter deployment, it was alleged that multiple ivc filter struts perforated the patient¿s inferior vena cava.The device has not been removed and there were no reported attempts made to retrieve the filter.The status of the patient is unknown.
 
Event Description
It was reported through the litigation process that a vena cava filter was placed in a patient after being diagnosed with deep vein thrombosis/pulmonary embolism.At some time post filter deployment, it was alleged that multiple ivc filter struts perforated the inferior vena cava.The device has not been removed and there were no reported attempts made to retrieve the filter.The status of the patient is unknown.
 
Manufacturer Narrative
H10: manufacturing review: a device history review (dhr) review could not be performed as the lot number is unknown.Investigation summary: the device was not returned for evaluation.Medical records were provided and reviewed.Computed tomography (ct) revealed the positive grade 2 perforation at most 3 mm and no significant tilt with coronal plane 6 degrees.Filter was located in superior l3 to superior l4 and it was unable to fully assess migration.Therefore, the investigation is confirmed for the perforation of the ivc.Based upon the available information, the definitive root cause is unknown.Labeling review: a review of product labeling documents (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, and unit label) showed that the product labeling is adequate.H10: g3,g4.H11: h6(device, method, conclusion).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G2 FILTER SYSTEM
Type of Device
VENA CAVA FILTER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key7602046
MDR Text Key111148240
Report Number2020394-2018-00868
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,health professional
Type of Report Initial,Followup
Report Date 06/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/14/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNKNOWN G2
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Date Manufacturer Received06/17/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age71 YR
-
-