• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-751NAS; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-751NAS; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-751NAS
Device Problems Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138); Device Inoperable (1663)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/25/2018
Event Type  malfunction  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
The customer reported via phone call that the insulin pump had a keypad anomaly.The buttons were not responsive.The screen was frozen.Customer's blood glucose level was not reported.Customer was advised to discontinue use of the device and revert to a backup plan.The customer was advised that the device would be replaced and agreed to return the product for analysis.
 
Manufacturer Narrative
No frozen screen, blank display or button error alarm noted.Unit time/clock moved.Unit had unresponsive all buttons due to unlocked lcd keypad connector.Unable to perform operating currents, self-test, unexpected restart error test and displacement test due to unresponsive button.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
530G INSULIN PUMP MMT-751NAS
Type of Device
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
MDR Report Key7604637
MDR Text Key111695255
Report Number3004209178-2018-82002
Device Sequence Number1
Product Code OZO
UDI-Device Identifier00643169503717
UDI-Public(01)00643169503717
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial,Followup
Report Date 09/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/15/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-751NAS
Device Catalogue NumberMMT-751NAS
Device Lot NumberA4751NASJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/15/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received08/23/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/28/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age54 YR
Patient Weight291
-
-