Model Number 9500-27
Device Problems Imprecision (1307); Use of Device Problem (1670)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/17/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
Dexcom was made aware on (b)(4) 2018 that on (b)(6) 2018, there was an inaccuracy between the continuous glucose monitor (cgm) and the blood glucose (bg) meter.The sensor was inserted into the abdomen on (b)(6) 2018.It was reported that the patient did not calibrate after the inaccuracy.No data was provided for evaluation.The complaint confirmation and probable cause could not be determined.Labeling indicates: if the difference between your sensor glucose reading and blood glucose value is greater than 20% of the blood glucose value for sensor glucose readings > 80 mg/dl or greater than 20 points for sensor glucose readings < 80 mg/dl, wash your hands and take another blood glucose measurement.If the difference between this second blood glucose measurement and the sensor is still greater than 20% for sensor glucose readings > 80 mg/dl or greater than 20 points for sensor glucose readings < 80 mg/dl, recalibrate your sensor using the second blood glucose value.The sensor glucose reading will correct over the next 15 minutes.No additional event or patient information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
Type of Device
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence dr.
san diego CA 92121
Manufacturer Contact
kipp durbin
6340 sequence dr.
san diego, CA 92121
8582000200
MDR Report Key7604922
MDR Text Key111177430
Report Number3004753838-2018-065114
Device Sequence Number1
Product Code MDS
UDI-Device Identifier00386270000019
UDI-Public00386270000019
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P120005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial
Report Date 05/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/15/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number9500-27
Device Catalogue NumberSTS-GL-011
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received05/21/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age58 YR
Patient Weight59
-
-