Model Number 20E
Device Problem Device Inoperable (1663)
Patient Problem No Patient Involvement (2645)
Event Date 05/01/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4). Physio-control evaluated the customer's device and verified the reported issue. Physio replaced the power supply assembly, which resolved the reported issue. After observing proper device operation through functional and performance testing, the device was returned to the customer for use. Physio-control further evaluated the removed power supply assembly and determined the cause of the reported issue was due to residue under filter component, designator fl10. This caused the device to cycle power.
 
Event Description
The customer contacted physio-control to report that the screen on their device was flashing on/off and they were unable to power the device on/off. The battery was removed to power the device off. This is indicative of a device locking up and as a result, defibrillation may not be possible, if it were necessary. There was no patient use associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameLIFEPAK® 20E DEFIBRILLATOR/MONITOR
Type of DeviceAUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
megan marseglia
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key7607507
MDR Text Key111297661
Report Number0003015876-2018-00950
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K130454
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation
Remedial Action Recall
Type of Report Initial
Report Date 06/15/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received06/15/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number20E
Device Catalogue Number99507-000133
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/11/2018
Is the Reporter a Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received05/30/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/17/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction Number3015876-11/07/2017-003C

-
-