Model Number 3660
Device Problem Device Inoperable (1663)
Patient Problem Inadequate Pain Relief (2388)
Event Date 05/18/2018
Event Type  malfunction  
Manufacturer Narrative
The device is included in the neuromodulation implantable pulse generator (ipg) inoperable when exposed to monopolar electrosurgery advisory notice issued by abbott on (b)(6) 2017.The manufacturer has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.The manufacturer defers to the patient's physician regarding medical history.
 
Event Description
It was reported the patient lost therapy and the ipg would no longer communicate with their patient controller after recent neck surgery.It was noted the patient's ipg was not put into surgery mode prior to having neck surgery.Troubleshooting was performed and therapy was restored.The issue is resolved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROCLAIM 5 ELITE
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer Contact
melissa nkematu
6901 preston road
plano, TX 75024
9723092520
MDR Report Key7607591
MDR Text Key111298655
Report Number1627487-2018-05485
Device Sequence Number1
Product Code LGW
UDI-Device Identifier05415067020192
UDI-Public05415067020192
Combination Product (y/n)N
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Remedial Action Other
Type of Report Initial
Report Date 06/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/15/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/28/2018
Device Model Number3660
Device Lot Number5474513
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received05/18/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/28/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number1627487-060217-001-C
Patient Sequence Number1
Patient Outcome(s) Other;
-
-