Model Number MMT-723LNAS
Device Problems Crack (1135); Improper or Incorrect Procedure or Method (2017)
Patient Problem Hyperglycemia (1905)
Event Date 05/21/2018
Event Type  Injury  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
The customer reported via phone call that they experienced high blood glucose and insulin pump was damaged.Customer¿s blood glucose was 455 mg/dl.The customer reported that there was crack around the seam of the plastic glass near the reservoir compartment.Troubleshooting was performed for damage and noticed the reservoir the lip was cracked and troubleshooting was not done for damage high blood glucose.Advised the pump will need to be replaced.Advised to discontinue use of the pump and revert to a backup plan.The insulin pump will be returned for analysis.
 
Manufacturer Narrative
The device passed rewind, basic occlusion, occlusion, prime/compromised force sensor system, excessive no delivery and displacement tests.The device had stripped battery cap coin slot (p).The reservoir tube lip was partially broken off but the test reservoir locked in place properly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723LNAS
Type of Device
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
MDR Report Key7608776
MDR Text Key111303866
Report Number3004209178-2018-82382
Device Sequence Number1
Product Code OYC
UDI-Device Identifier00643169513914
UDI-Public(01)00643169513914
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial,Followup
Report Date 09/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/18/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberMMT-723LNAS
Device Catalogue NumberMMT-723LNAS
Device Lot NumberA5723LNASJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/06/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received08/16/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/02/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age51 YR
Patient Weight254
-
-