ROCHE DIAGNOSTICS ELECSYS TSH ASSAY; RADIOIMMUNOASSAY, THYROID-STIMULATING HORMONE Back to Search Results
Catalog Number 07028091190
Device Problem Incorrect Or Inadequate Test Results (2456)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/22/2018
Event Type  malfunction  
Manufacturer Narrative
This event occurred in (b)(6).(b)(4).
 
Event Description
The customer questioned results from 1 patient sample tested for elecsys tsh (tsh), elecsys ft4 ii (ft4 ii) and elecsys ft3 iii (ft3 iii) on a cobas e801 module.The sample was submitted for investigation and discrepant tsh results were identified between the customer's e801 module, the centaur method, a cobas 8000 e 602 module used at the investigation site and a cobas e 411 immunoassay analyzer.It is not known if incorrect results were reported outside of the laboratory.Refer to attached data for the patient results.The patient sample was also treated and underwent polyethylene glycol (peg) testing.Refer to attached data for these results.There was no allegation that an adverse event occurred.The serial number for the customer's e801 module was not provided.The e602 serial number was 1286-07.The e411 analyzer serial number was 1368-02.The tsh reagent lot number used with the e602 module was 294897 with an expiration date of 31-aug-2018.The tsh reagent lot number used with the e411 analyzer was 291617 with an expiration date of 31-jul-2018.Calibration and qc at the investigation site were acceptable.From the data provided, a general reagent issue can be excluded.Assays from different manufacturers can generate different results.The investigation was unable to find a definitive root cause.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS TSH ASSAY
Type of Device
RADIOIMMUNOASSAY, THYROID-STIMULATING HORMONE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key7609261
MDR Text Key112157474
Report Number1823260-2018-01943
Device Sequence Number1
Product Code JLW
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K961491
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Health Professional
Type of Report Initial
Report Date 06/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/18/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number07028091190
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/31/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-