The customer questioned results from 1 patient sample tested for elecsys tsh (tsh), elecsys ft4 ii (ft4 ii) and elecsys ft3 iii (ft3 iii) on a cobas e801 module.The sample was submitted for investigation and discrepant tsh results were identified between the customer's e801 module, the centaur method, a cobas 8000 e 602 module used at the investigation site and a cobas e 411 immunoassay analyzer.It is not known if incorrect results were reported outside of the laboratory.Refer to attached data for the patient results.The patient sample was also treated and underwent polyethylene glycol (peg) testing.Refer to attached data for these results.There was no allegation that an adverse event occurred.The serial number for the customer's e801 module was not provided.The e602 serial number was 1286-07.The e411 analyzer serial number was 1368-02.The tsh reagent lot number used with the e602 module was 294897 with an expiration date of 31-aug-2018.The tsh reagent lot number used with the e411 analyzer was 291617 with an expiration date of 31-jul-2018.Calibration and qc at the investigation site were acceptable.From the data provided, a general reagent issue can be excluded.Assays from different manufacturers can generate different results.The investigation was unable to find a definitive root cause.
|