• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEXCOM, INC. DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEXCOM, INC. DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 9438-06
Device Problem Low Battery (2584)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/25/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
Dexcom was made aware on (b)(4) 2018, that on (b)(6) 2018, the smart device displayed an error message for low transmitter battery.Data was provided for evaluation.Data review could confirm the reported event of low transmitter battery.Additionally; the transmitter has been received for evaluation.An external visual inspection was performed and the passed.A voltage test was performed and the transmitter failed as it had no voltage.Review of the log data was performed and transmitter failure was observed.The reported issue of low transmitter battery was not confirmed via product testing.This complaint is reportable based on the finding of transmitter failure via log data review.A probable cause is low transmitter battery voltage.No additional patient or event information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM
Type of Device
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence dr.
san diego CA 92121
Manufacturer Contact
kipp durbin
6340 sequence dr.
san diego, CA 92121
8582000200
MDR Report Key7609669
MDR Text Key111496220
Report Number3004753838-2018-065416
Device Sequence Number1
Product Code MDS
UDI-Device Identifier10386270000221
UDI-Public10386270000221
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P120005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial
Report Date 05/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/18/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number9438-06
Device Catalogue NumberSTT-GF-001
Device Lot Number6000317
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/14/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received05/25/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/15/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age28 YR
-
-