Results: the neuron max distal tip did not show evidence of damage.The neuron max was fractured approximately 3.0 cm from the hub and was kinked approximately 83.0 cm from the hub.Conclusions: evaluation of the returned neuron max revealed a device with an undamaged distal atraumatic tip.The neuron max tip was compared with a demonstration device and were indistinguishable from each other.Further evaluation revealed that the neuron max was kinked and fractured.If the neuron max is forcefully manipulated at extreme angles when removed from packaging or otherwise mishandled during preparation, the device may become fractured.Based on the return condition, the kink was likely incidental to the reported failure and may have occurred due to the packaging for return to penumbra.Penumbra catheters are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
During preparation for a medical procedure, the hospital staff found that a neuron max 6f 088 long sheath (neuron max) had a tear at the distal tip upon removal from the packaging.The damaged neuron max was found prior to use, and therefore, was not used in the procedure.The procedure was completed using another neuron max.
|