• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES PRODUCTS LLC SMALL BATTERY DRIVE II; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES PRODUCTS LLC SMALL BATTERY DRIVE II; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT Back to Search Results
Catalog Number 532.110
Device Problem Detachment Of Device Component (1104)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/10/2018
Event Type  malfunction  
Manufacturer Narrative
The actual device was returned for evaluation.The device was evaluated and the reported condition was confirmed.A visual and functional assessment was performed which found that the device was not working at all.It was determined that the device failed the safety check for off/oscillation/on switch mode function.During service/repair, it was observed that the motor was not rotating and was corroded, and the bearings and gear parts were also corroded.It was observed that the electronic control unit (ecu) function was damaged and the cover plate had come off.The assignable root cause was determined to be due to improper cleaning and normal wear.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Event Description
It was reported that during pre-surgery, it was discovered that the small battery drive device was not working.During in-house engineering evaluation, it was observed that the electronic control unit (ecu) was damaged and the cover plate had come off.It was not reported if there were any delays to the surgical procedure or if a spare device was available.There was no patient involvement reported.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMALL BATTERY DRIVE II
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
OBERDORF SYNTHES PRODUKTIONS GMBH
eimattstrasse 3
oberdorf 4436
SZ   4436
Manufacturer Contact
michael cote
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key7613085
MDR Text Key111443014
Report Number8030965-2018-54401
Device Sequence Number1
Product Code HWE
UDI-Device Identifier7611819408357
UDI-Public(01)7611819408357(11)141209(10)1605
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 04/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/19/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number532.110
Device Lot Number1605
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/25/2018
Date Manufacturer Received06/06/2018
Date Device Manufactured12/09/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-