DEPUY SYNTHES PRODUCTS LLC SMALL BATTERY DRIVE II; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
|
Back to Search Results |
|
Catalog Number 532.110 |
Device Problem
Detachment Of Device Component (1104)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 04/10/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
The actual device was returned for evaluation.The device was evaluated and the reported condition was confirmed.A visual and functional assessment was performed which found that the device was not working at all.It was determined that the device failed the safety check for off/oscillation/on switch mode function.During service/repair, it was observed that the motor was not rotating and was corroded, and the bearings and gear parts were also corroded.It was observed that the electronic control unit (ecu) function was damaged and the cover plate had come off.The assignable root cause was determined to be due to improper cleaning and normal wear.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Event Description
|
It was reported that during pre-surgery, it was discovered that the small battery drive device was not working.During in-house engineering evaluation, it was observed that the electronic control unit (ecu) was damaged and the cover plate had come off.It was not reported if there were any delays to the surgical procedure or if a spare device was available.There was no patient involvement reported.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Search Alerts/Recalls
|
|
|