Model Number S7
Device Problems Device Stops Intermittently (1599); Communication or Transmission Problem (2896); Device Operates Differently Than Expected (2913)
Patient Problem No Patient Involvement (2645)
Event Date 05/23/2018
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.A medtronic representative went to the site to test the equipment.Testing revealed that the failure could not be found however a checking the software of the ndi toolbox configuration utility revealed errors in the positioning sensor unit (psu) and polaris spectra system control unit (scu).No resolution has been confirmed yet.A software investigation analysis was initiated to determine the probable cause of the issue through log analysis.Analysis found that the scu was having communication issues and psu having intermittent voltage issues with battery, sensor, illuminator.
 
Event Description
Medtronic received information regarding a navigation device.It was reported that the camera was restarting and not tracking instruments.Restarting the device resolved the did not resolve the issues but the camera becomes normal after powering down for a while.The reported issue had been occurring intermittently prior to reporting the complaint.There was no patient present when this issue was observed.
 
Manufacturer Narrative
A software investigation analysis was initiated to determine the probable cause of the reported issue.Analysis found that the software functioned as designed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S7 SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
prashanth gali
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
7208902515
MDR Report Key7614317
MDR Text Key111496309
Report Number1723170-2018-02859
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00643169505742
UDI-Public00643169505742
Combination Product (y/n)N
Reporter Country CodeCN
PMA/PMN Number
K050438
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
Report Date 08/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/19/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberS7
Device Catalogue Number9733858
Is the Reporter a Health Professional? No
Date Manufacturer Received07/24/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/17/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-