Model Number MODEL 100
Device Problems Device Issue (2379); Device Operates Differently Than Expected (2913); Device Operational Issue (2914)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/21/2018
Event Type  malfunction  
Manufacturer Narrative
The report of an unreadable autopulse platform (sn (b)(4)) user control panel was confirmed during functional testing; the root cause was due to a damaged user control panel lcd.As part of routine service during testing, the platform was examined and found physical damage on the top cover and the drive shaft exhibits binding and resistance.This observation is unrelated to the reported event.After replacement of the user control panel lcd and the damaged parts, the platform was functionally tested and operated as intended.The platform passed all testing specifications with no further issue observed.Historical records were reviewed for service information related to the reported complaint and there was no similar complaint reported for autopulse with serial number (b)(4).
 
Event Description
During routine inspection, the backlight of the autopulse platform (sn (b)(4)) user control panel did not illuminate.No patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE® RESUSCITATION MODEL 100
Type of Device
CARDIAC CHEST COMPRESSOR
Manufacturer (Section D)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer (Section G)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer Contact
kimthoa nguyen
2000 ringwood ave,
san jose, CA 95131
4084192922
MDR Report Key7617141
MDR Text Key111715426
Report Number3010617000-2018-00655
Device Sequence Number1
Product Code DRM
UDI-Device Identifier00849111000512
UDI-Public00849111000512
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112998
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial
Report Date 06/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/19/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 100
Device Catalogue Number8700-0730-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/24/2018
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received05/21/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/07/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-