Model Number N140 |
Device Problems
High impedance (1291); Connection Problem (2900); Device Contamination With Biological Material (2908)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 04/25/2018 |
Event Type
Injury
|
Manufacturer Narrative
|
At this time, the product has not been returned.If the product is returned, analysis will be performed and this report would be updated at that time.
|
|
Event Description
|
Boston scientific received information that this cardiac resynchronization therapy defibrillator (crt-d) and right ventricular (rv) lead had high out of range shock lead impedances greater than 125 ohms.The lead was explanted and replaced, and the device was explanted due to contamination and replaced.Additional information was requested to clarify the contamination.No adverse patient effects were reported.
|
|
Event Description
|
Additional information was received that further explained the report of device contamination.The lead was damaged with electrocautery during explant, causing the lead to break and get stuck in the header of the device.The device was not able to be reused, and was explanted.
|
|