Model Number N140
Device Problems High impedance (1291); Connection Problem (2900); Device Contamination With Biological Material (2908)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/25/2018
Event Type  Injury  
Manufacturer Narrative
At this time, the product has not been returned.If the product is returned, analysis will be performed and this report would be updated at that time.
 
Event Description
Boston scientific received information that this cardiac resynchronization therapy defibrillator (crt-d) and right ventricular (rv) lead had high out of range shock lead impedances greater than 125 ohms.The lead was explanted and replaced, and the device was explanted due to contamination and replaced.Additional information was requested to clarify the contamination.No adverse patient effects were reported.
 
Event Description
Additional information was received that further explained the report of device contamination.The lead was damaged with electrocautery during explant, causing the lead to break and get stuck in the header of the device.The device was not able to be reused, and was explanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENERGEN
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer (Section G)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer Contact
tim degroot
4100 hamline ave. n
st. paul, MN 
6515826168
MDR Report Key7617268
MDR Text Key111610998
Report Number2124215-2018-09564
Device Sequence Number1
Product Code LWP
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/19/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date08/02/2013
Device Model NumberN140
Other Device ID NumberENERGEN CRT-D
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Manufacturer Received04/25/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/06/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
Patient Age58 YR
-
-