• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GUIDANT CRM CLONMEL IRELAND ACCOLADE; IMPLANTABLE PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GUIDANT CRM CLONMEL IRELAND ACCOLADE; IMPLANTABLE PULSE GENERATOR Back to Search Results
Model Number L331
Device Problems Failure to Capture (1081); High impedance (1291); Device Contamination With Biological Material (2908)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/04/2018
Event Type  Injury  
Manufacturer Narrative
The product is expected to be returned for analysis.This report will be updated upon return and completion of analysis.
 
Event Description
Boston scientific received information that, shortly after implant, this pacemaker exhibited loss of capture and recorded a lead safety switch due to high out of range pacing impedances on the right ventricular (rv) channel.A revision procedure was performed and blood contamination was noted in the device header.The rv lead was tested via a pacing system analyzer (psa) with good parameters but the measurements remained high when reconnected to this pacemaker.This pacemaker was removed and replaced.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCOLADE
Type of Device
IMPLANTABLE PULSE GENERATOR
Manufacturer (Section D)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer (Section G)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer Contact
tim degroot
4100 hamline ave. n
st. paul, MN 
6515826168
MDR Report Key7618441
MDR Text Key111630682
Report Number2124215-2018-08910
Device Sequence Number1
Product Code LWP
Combination Product (y/n)N
Reporter Country CodeCA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Physician
Type of Report Initial
Report Date 05/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/20/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/22/2018
Device Model NumberL331
Other Device ID NumberACCOLADE MRI
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Manufacturer Received05/04/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/22/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
L331
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
-
-