(b)(4).
Evaluation summary: the device returned for analysis.
The complaint investigation determined the reported difficulty was related to manufacturing issues associated with the protective sheath.
On march 14, 2017, abbott vascular decided to initiate a voluntary field action for some sizes and lots of the nc trek family of dilatation catheters for difficulty to remove sheath which may lead to inflation or deflation issues.
Abbott vascular performed a comprehensive investigation which included device analysis, manufacturing evaluation and trend analysis.
The root cause identification was complicated by the fact that users were describing multiple symptoms when reporting the complaints.
To date, the frequency of worldwide reported events for difficulties removing the protective balloon sheath, inflation and deflation has reached an actionable limit, thus abbott vascular communicated the voluntary field action to the fda on march 17, 2017 [medwatch # 2024168-2017-02310].
Corrective action has been implemented per site operating procedures.
The product will continue to be trended.
|
It was reported that prior to use, during stylet removal, the protective sheath was difficult to remove from the 3.
5x15 mm nc trek rx balloon dilatation catheter (bdc), resulting in a balloon detachment.
The nc trek bdc was not used for the procedure.
There was no patient involvement.
An unspecified balloon dilatation catheter was used to complete the procedure.
There was no clinically significant delay during the procedure.
No additional information was provided.
|