Medtronic received information regarding a navigation system being used during a functional endoscopic sinus surgery (fess).
It was reported that intra-operatively, after getting a passing registration, the site was not able to move forward and click the ¿verify accuracy¿ button even though it was highlighted blue.
The site had passing registrations of 1.
9 and 2.
2 in previous registrations but couldn¿t proceed forward in the software with either registration.
A manufacturer representative logged out and then logged back in, and all the previous registrations were red status and they couldn¿t select any.
The representative tried to delete the patient and reload to perform another registration with a blank slate, but the site proceeded without navigation.
Surgical time was extended by less than one hour and there was no reported patient impact.
|