Catalog Number RF310F
Device Problems Extrusion (2934); Patient Device Interaction Problem (4001)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/05/2014
Event Type  malfunction  
Manufacturer Narrative
Manufacturing review: the device history records have been reviewed with special attention to the raw materials, subassemblies, manufacturing process and quality control testing.This lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only event reported to date for this lot number and failure mode.Medical record review: patient with a history of left femoral deep vein thrombosis had a vena cava filter deployed infrarenal.Placement was confirmed with x-ray.The patient tolerated the procedure well and moved to the floor in stable condition.Investigation summary: the device was not returned for evaluation.Images were not provided for review.Medical records were provided and reviewed.There was no specific deficiency alleged in the provided medical records.Therefore, the investigation is inconclusive for perforation as no objective evidence has been provided to confirm any alleged deficiency with the filter.Based upon the available information, the definitive root cause is unknown.Labeling review: a review of product labeling documents (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, and unit label) showed that the product labeling is adequate.(expiry date: 08/2014); (08/2011).The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event Description
It was reported through the litigation process that a vena cava filter was placed in a patient after being diagnosed with deep vein thrombosis/pulmonary embolism.At some time post filter deployment, it was alleged that the filter perforated the vena cava with struts of the filter abutting the duodenum, l4 vertebral wall, and aorta.The device has not been removed and there were no reported attempts made to retrieve the filter.The status of the patient is unknown.
 
Event Description
It was reported through the litigation process that a vena cava filter was placed in a patient after being diagnosed with deep vein thrombosis/pulmonary embolism.At some time post filter deployment, it was alleged that the filter perforated the patient¿s vena cava with struts of the filter abutting the patient¿s duodenum, l4 vertebral wall, and aorta.The device has not been removed and there were no reported attempts made to retrieve the filter.The status of the patient is unknown.
 
Manufacturer Narrative
H10: manufacturing review: the device history records have been reviewed with special attention to the raw materials, subassemblies, manufacturing process, and quality control testing.This lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the device was not returned for evaluation.The medical records included images.The image review was documented in the medical records.Medical records were provided and reviewed.Approximately two years later a computed tomography shows that the inferior vena cava filter was at sup l3 to mid l4, there was no evidence of migration and tilt and there was grade 3 perforation evident with right lateral strut to the duodenum, posterior strut at superior l4 and let lateral strut at aorta.Therefore, the investigation can be confirmed for perforation of the ivc.Based upon the available information, the definitive root cause is unknown.Labeling review: a review of product labeling documents (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, and unit label) showed that the product labeling is adequate.H10: b6,d4 (expiry date:08/2014).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G2 FILTER SYSTEM - FEMORAL
Type of Device
VENA CAVA FILTER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key7622180
MDR Text Key111915831
Report Number2020394-2018-00929
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
PMA/PMN Number
K062887
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,health professional
Type of Report Initial,Followup
Report Date 06/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/20/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberRF310F
Device Lot NumberGFVG2310
Was Device Available for Evaluation? No
Date Manufacturer Received06/17/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age57 YR
-
-