• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - COSTA RICA (COYOL) DISPOSABLE SUTURE PLACEMENT SYSTEM; LAPAROSCOPE, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - COSTA RICA (COYOL) DISPOSABLE SUTURE PLACEMENT SYSTEM; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number UNK55
Device Problem Detachment Of Device Component (1104)
Patient Problem Device Embedded In Tissue or Plaque (3165)
Event Date 04/24/2018
Event Type  Injury  
Manufacturer Narrative
The complainant was unable to provide the suspect device lot number; therefore, the lot expiration and device manufacture dates are unknown.Device component code is related to device problem code for the problem of needle detachment.The complaint device is not expected to be returned for evaluation; therefore, a failure analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.The voluntary user medwatch number is mw5077142.
 
Event Description
It was reported to boston scientific corporation that a capio device was used during a procedure performed on (b)(6) 2018 according to the complainant, during the procedure, upon deployment of the capio device on the patient's right side, the dart detached and got stuck in the periosteum of the pubic bone.They then opened the area and the retropubic space but they were unable to retrieve the dart.An op x-ray on the pelvic floor was performed to locate the dart and the dart was shown around the right side of the pubic bone.Several dissection performed in an attempt to retrieve the dart, but they were still unable to retrieve the dart.All other information is unknown.Should additional relevant details become available; a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISPOSABLE SUTURE PLACEMENT SYSTEM
Type of Device
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
BOSTON SCIENTIFIC - COSTA RICA (COYOL)
2546 first street
propark free zone
alajuela
CS 
Manufacturer (Section G)
BOSTON SCIENTIFIC - COSTA RICA (COYOL)
2546 first street
propark free zone
alajuela
CS  
Manufacturer Contact
nancy cutino
100 boston scientific way
marlborough, MA 01752
5086834000
MDR Report Key7623603
MDR Text Key111819958
Report Number3005099803-2018-01879
Device Sequence Number1
Product Code GCJ
Combination Product (y/n)N
PMA/PMN Number
K932553
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial
Report Date 05/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/20/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUNK55
Was Device Available for Evaluation? No
Date Manufacturer Received05/29/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-